Please use this identifier to cite or link to this item: http://dspace.utpl.edu.ec/handle/123456789/32264
Full metadata record
DC FieldValueLanguage
dc.contributor.advisorEspinosa Tituana, Lupe Carolina.es_ES
dc.contributor.authorCabrera Briceño, María Paulaes_ES
dc.date.accessioned2022-09-30T15:04:19Z-
dc.date.available2022-09-30T15:04:19Z-
dc.date.issued2022es_ES
dc.identifier.citationCabrera Briceño, M. P. Espinosa Tituana, L. C. (2022) Validación de un método de cromatografía líquida de alta eficacia (HPLC) para la valoración de crisaborol en formulaciones tópicas [Tesis de Grado, Universidad Técnica Particular de Loja]. Repositorio Institucional. https://dspace.utpl.edu.ec/handle/123456789/32264es_ES
dc.identifier.otherCobarc: 1355989es_ES
dc.identifier.urihttps://bibliotecautpl.utpl.edu.ec/cgi-bin/abnetclwo?ACC=DOSEARCH&xsqf99=131279.TITN.es_ES
dc.descriptionResumen: En el presente trabajo se describe la optimización de la técnica analítica de cromatografía líquida de alta resolución (HPLC)mediante el método de elución isocrático para la valoración de crisaborol (CRB), un nuevo inhibidor tópico no esteroideo de la fosfodiesterasa-4. La separación del analito se logró por medio de una columna Kromasil tipo C18 (4.6 x250 mm, 5 μm), utilizando agua: acetonitrilo (30:70 v/v) como fase móvil, a un caudal de 0,75 ml/min a temperatura ambiente y detección de PDA a 285 nm. Además, se presenta la validación, evaluándose los parámetros de linealidad, especificidad, sensibilidad, exactitud y precisión bajo la normativa Q2-R1 de las Conferencias Internacionales de Armonización de requerimientos técnicos para el registro de medicamentos de uso humanos (ICH).es_ES
dc.description.abstractAbstract: The present work describes the optimization of a high-performanceliquid chromatography (HPLC) method, in isocratic elution mode for the titration of chrysaborole (CRB), a new non-steroidal topical phosphodiesterase-4 inhibitor; the separation of the analyte was achieved by means of a Kromasil column type C18 (4. 6 x250mm, 5 μm), using water: acetonitrile (30:70 v/v) as mobile phase, at a flow rate of 0.75 ml/min at room temperature and PDA detection at 285 nm.In addition, the validation is presented, evaluating the parameters of linearity, specificity, sensitivity, accuracy and pressure under the Q2-R1 standard of the International Conferences on Harmonization of Technical Requirements for Registration of Medicinal Products for Human Use (ICH)es_ES
dc.language.isospaes_ES
dc.subjectEcuador.es_ES
dc.subjectTesis digital.es_ES
dc.titleValidación de un método de cromatografía líquida de alta eficacia (HPLC) para la valoración de crisaborol en formulaciones tópicases_ES
dc.typebachelorThesises_ES
Appears in Collections:Titulación de Bioquimica y Farmacia



Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.